Scanning probe detector channel descriptions

From CNM Wiki
Jump to navigation Jump to search

Back to Scanning Probe Diffraction Quick Reference

Typical definitions of 26ID scan detector channels

Detector Channel number EPICS Process variable Description
D01 26idsclr1:scaler1.S1 I0 - Integrated Ion chamber current after NES slits (change current amplification via yellow SR570 tab on main beamline screen)
D02 26idcXMAP:mca5.R1 Bi - Fluorescence channel from outboard detector, summed ROI#1 - find and modify with MCA software from User Tools tab on main beamline screen
D03 26idcXMAP:mca5.R2 Ti - Fluorescence channel from outboard detector, summed ROI#2 - find and modify with MCA software from User Tools tab on main beamline screen
D04 26idcXMAP:mca5.R5 Fe - Fluorescence channel from outboard detector, summed ROI#5 - find and modify with MCA software from User Tools tab on main beamline screen
D05 26id:ccd.R01.Average Integrated Intensity/Num Pixels for CCD ROI#1 - ROI#1 defines the total active chip area to be saved, so this is the total CCD intensity per pixel
D06 26id:ccd.R02.Average Integrated Intensity/Num Pixels for CCD ROI#2 - usually tightly defined around the outgoing beam - Contrast here is related to diffracting volume
D07 26id:ccd.R02.XCentroid X center of mass of intensity in CCD ROI#2 - This is related to lattice strain and rotation in the diffraction plane, scales away from center of ROI2 with signal/dark current
D08 26id:ccd.R02.YCentroid Y center of mass of intensity in CCD ROI#2 - This is related to lattice rotation out of the diffraction plane, scales away from center of ROI2 with signal/dark current
D09 26idcDET:userCalc3.VAL R02.Average - R03.Average, Background subtracted average intensity - ROI#3 typically set around an unused area, this is useful for viewing scans in progress with low contrast
D10 26idcDET:userCalc6.VAL CCD Image number - If you are using the CCD DO NOT MODIFY THIS CHANNEL as it is necessary for Matlab analysis routines
D11 26idctsrc:DMM1:Ch1_raw.VAL Channel 1 - C Hutch digital multimeter - FOMX temperature
D12 26idctsrc:DMM1:Ch2_raw.VAL Channel 2 - C Hutch digital multimeter - FOMY temperature
D13 26idctsrc:DMM1:Ch3_raw.VAL Channel 3 - C Hutch digital multimeter - FOMZ temperature
D14 26idctsrc:DMM1:Ch4_raw.VAL Channel 4 - C Hutch digital multimeter - SAMX temperature
D15 26idctsrc:DMM1:Ch5_raw.VAL Channel 5 - C Hutch digital multimeter - SAMY temperature
D16 26idctsrc:DMM1:Ch6_raw.VAL Channel 6 - C Hutch digital multimeter - SAMZ temperature
D17 26idctsrc:DMM1:Ch7_raw.VAL Channel 7 - C Hutch digital multimeter - SAM Theta temperature
D18 26idpvc:userStringCalc3.VAL Time of day - Units are fractional hours (ie 22.5 = 10:30pm) - use min and max of a linear scan to accurately estimate 2D scanning times, discontinuity in this channel indicates a hung scan
D19 - -
D20 - -
D21 B26ID-CG-DCM-01:PFEED Feed pressure of LN2 cryocooler internal circuit - spikes at refill, adjust fill cycle levels and remove ice from vent line to minimize this
D22 B26ID-CG-DCM-01:MNLEV Cryocooler main vessel LN2 level - direct measure of LN2 fill cycle, dewar currently is kept between 80-60% full
D23 26id:BLEPS:MIRROR TEMP 01:RBV First mirror temperature - RTD readout - useful to monitor equilibration after a beam dump
D24 26id:BLEPS:MIRROR TEMP 02:RBV Second mirror temperature
D25 26id:BLEPS:XTAL_TEMP_01:RBV DCM first crystal temperature - typically -234 deg F
D26 26id:BLEPS:XTAL_TEMP_02:RBV DCM second crystal temperature - typically -171 deg F
D27 26idctsrc:bpmds:current:total Alternate I0 - integrated intensity on all four quadrant diodes at the front of C Hutch - mono beam intensity measurement that does not depend on NES slit settings or exposure times
D28 26idctsrc:bpmds:pos:x Approximate X (horizontal) position of mono beam in microns at the front of C hutch - related to mirror piezo and DCM chi2 stability
D29 26idctsrc:bpmds:pos:y Approximate Y (vertical) position of mono beam in microns measured at the front of C hutch - primarily related to DCM theta2 stability
D30 S:SRcurrentAI Alternate I0 - APS storage ring current value, useful for normalizing long scan data when the ring is not in top-up mode - unexpected changes of this value indicate orbit/steering loss or beam dump
D31 26idbDCM:userCalc1.VAL Fe fluorescence - Integrated signal from the four elements of the inboard detector (26idcXMAPmca1 - 26idcXMAPmca4) - ROI#8
D32 26idbDCM:userCalc2.VAL Cu fluorescence - Integrated signal from the four elements of the inboard detector (26idcXMAPmca1 - 26idcXMAPmca4) - ROI#11
D33 26idbDCM:userCalc3.VAL Zn fluorescence - Integrated signal from the four elements of the inboard detector (26idcXMAPmca1 - 26idcXMAPmca4) - ROI#13
D34 26idbDCM:userCalc4.VAL Co fluorescence - Integrated signal from the four elements of the inboard detector (26idcXMAPmca1 - 26idcXMAPmca4) - ROI#9
D35 26idbDCM:userCalc5.VAL Ti fluorescence - Integrated signal from the four elements of the inboard detector (26idcXMAPmca1 - 26idcXMAPmca4) - ROI#4
D36 26idbDCM:userCalc6.VAL Ca fluorescence - Integrated signal from the four elements of the inboard detector (26idcXMAPmca1 - 26idcXMAPmca4) - ROI#3
D37 26idbDCM:userCalc7.VAL Mn fluorescence - Integrated signal from the four elements of the inboard detector (26idcXMAPmca1 - 26idcXMAPmca4) - ROI#7
D38 26idbDCM:userCalc8.VAL Si fluorescence - Integrated signal from the four elements of the inboard detector (26idcXMAPmca1 - 26idcXMAPmca4) - ROI#1
D39 26idbDCM:userCalc9.VAL Al fluorescence - Integrated signal from the four elements of the inboard detector (26idcXMAPmca1 - 26idcXMAPmca4) - ROI#0
D40 26idbDCM:userCalc10.VAL K fluorescence - Integrated signal from the four elements of the inboard detector (26idcXMAPmca1 - 26idcXMAPmca4) - ROI#2
D41 26idbPBS:userCalc5.VAL V fluorescence - Integrated signal from the four elements of the inboard detector (26idcXMAPmca1 - 26idcXMAPmca4) - ROI#5
D42 26idbPBS:userCalc4.VAL Full spectra - Integrated signal from the four elements of the inboard detector (26idcXMAPmca1 - 26idcXMAPmca4) over all energies- ROI#15
D43 26idbPBS:userCalc6.VAL P fluorescence - Integrated signal from the four elements of the inboard detector (26idcXMAPmca1 - 26idcXMAPmca4) - ROI#14
D44 26idbDCM:userCalc8.VAL Au fluorescence - Integrated signal from the four elements of the inboard detector (26idcXMAPmca1 - 26idcXMAPmca4) - ROI#10
D45 - -
D46 26id:ccd.Milliseconds CCD exposure time - useful for scan to scan normalization
D47 26idcXMAP:PresetReal Fluorescence detector exposure time
D48 - -
D49 - -
D50 - -
D51 26idcnpi:m10.RBV Focus X readback value - these next ten detectors are the absolute positions of the NPI scanning axes this should allow you to fully recreate the NPI scanning position at any data point
D52 26idcnpi:m11.RBV Focus Y readback value
D53 26idcnpi:m12.RBV Focus Z readback value
D54 26idcnpi:m16.RBV Sample X readback value
D55 26idcnpi:m17.RBV Sample Y readback value
D56 26idcnpi:m18.RBV Sample Z readback value
D57 26idcnpi:m20.RBV Sample Theta readback value
D58 26idcDET:base:Theta_d Two theta detector position
D59 26idcnpi:m34.RBV Piezo X readback value - differential Focus-Sample X position measured by the laser reference frame
D60 26idcnpi:m35.RBV Piezo Y readback value - differential Focus-Sample Y position measured by the laser reference frame