Some more notes about flashing digitizers or triggers:

  1. Log into dgs account on dgs1
  2. Get digitizer.bin from directory /Digitizer/MAIN_FPGA/Work11_DGS
  3. copy to /home/dgs/tmadden/DGSDigFirmware or /home/dgs/tmadden/DGSTrigFirmware
  4. cd /home/dgs/tmadden/swWork/workspace/epicsClient/src
  5. issue following: java -classpath jca-2.3.5.jar:caj-1.1.9.jar:fpgasender.jar plotControl
  6. copy and paste below

java -classpath jca-2.3.5.jar:caj-1.1.9.jar:fpgasender.jar plotControl
epics.epics_init();
var fn0=new String("/home/dgs/tmadden/swWork/workspace/epicsClient/src/asynRecords.txt");
var fn1=new String("/home/dgs/tmadden/retfile.bin");
epics.connectPVs(fn0);
var digware=new String("/home/dgs/tmadden/DGSDigFirmware/MSTR_digitizer_20200722.bin");
var slvware=new String("/home/dgs/tmadden/DGSDigFirmware/SLAVE_digitizer_20200602.bin");
var mastware=new String("/home/dgs/tmadden/DGSTrigFirmware/trigger_top_20160626.bin");
var routware=new String("/home/dgs/tmadden/DGSTrigFirmware/router_top_20160601.bin");

NOTEL Need to define add current name of master, slave digitizer bin files and trigger & router bin file names