The detectors in Gammasphere are [[High Purity Germanium (HPGe)|high purity germanium]] detectors sensitive to x- and gamma-rays from a few dozen keV to ~10MeV energy.  All Gammasphere detectors are Compton shielded using bismuth germanate (BGO) shields.  Six planar BGO detectors form a hexagonal shield array around the HPGe detector sides and a 7th BGO shield behind (at farther radius from the interaction point) is used as a "back plug".  The germanium detector itself is of the coaxial type.  Incident gamma-rays typically deposit charge in the germanium that is read out through the center contact.
The detectors in Gammasphere are [[High Purity Germanium (HPGe)|high purity germanium]] detectors sensitive to x- and gamma-rays from a few dozen keV to ~10MeV energy.  All Gammasphere detectors are Compton shielded using bismuth germanate (BGO) shields.  Six planar BGO detectors form a hexagonal shield array around the HPGe detector sides and a 7th BGO shield behind (at farther radius from the interaction point) is used as a "back plug".  The germanium detector itself is of the coaxial type.  Incident gamma-rays typically deposit charge in the germanium that is read out through the center contact.


Should the gamma-ray scatter within the Ge portion of the detector the scatter will normally be detected by interaction with the BGO shields.  Thus we define a 'clean' event as one that excites the Ge but none of the surrounding BGO of a given detector and a 'dirty' event as one in which both the Ge and BGO light up.  Within the digitizing data acquisition system, the sum of all 7 BGO signals is applied to one digitizer channel and the Ge center contact is digitized by another channel in the same digitizer.  A discriminator threshold is applied to both signals and the Router trigger module looks at the time relationship between Ge discriminator bits and the associated BGO Sum discriminator bitsIf the two discriminator bits are close enough in time the event is marked as 'dirty'.
The data that is collected in each detector goes through the [[DAQ System]], the data acquisition system which presents results to the user.   


=== Segmented Detectors ===
=== Segmented Detectors ===

Revision as of 19:08, February 15, 2023

This image is interactive. Click text to go to the link.

The Slope BoxThe Slope Box ExtensionTest Image Map
About this image

The detectors in Gammasphere are high purity germanium detectors sensitive to x- and gamma-rays from a few dozen keV to ~10MeV energy. All Gammasphere detectors are Compton shielded using bismuth germanate (BGO) shields. Six planar BGO detectors form a hexagonal shield array around the HPGe detector sides and a 7th BGO shield behind (at farther radius from the interaction point) is used as a "back plug". The germanium detector itself is of the coaxial type. Incident gamma-rays typically deposit charge in the germanium that is read out through the center contact.

The data that is collected in each detector goes through the DAQ System, the data acquisition system which presents results to the user.

Segmented Detectors

Many of the Ge detectors in Gammasphere are segmented. In these the Ge portion of the detector is split in half resulting in three Ge contacts (Center, Side A, Side B). Only one of the side connections is brought out to the digitizers as the charge that would be in the other side is given by subtraction of the measured side from the center contact.

Non-Segmented Detectors

Non-Segmented detectors are the same as segmented ones, but the Ge inside of the detector isn't split, resulting in two Ge contacts instead of three.