Run procedures: Difference between revisions
Tlauritsen (talk | contribs) |
(Removed incorrect commentary.) |
||
Line 4: | Line 4: | ||
* Start/Stop - this is self explanatory. Pushing to start begins taking data. | * Start/Stop - this is self explanatory. Pushing to start begins taking data. | ||
* NoSave/Save - If set to Save, IOC buffers fill with data | * NoSave/Save - If set to Save, IOC buffers fill with data. | ||
Line 45: | Line 45: | ||
and then stop the receivers : type Ctrl-C on all the receivers (IOC1, OIC2 ....) | and then stop the receivers : type Ctrl-C on all the receivers (IOC1, OIC2 ....) | ||
==Linking two or more systems== | ==Linking two or more systems== | ||
Line 72: | Line 69: | ||
'''Note:''' Clock Source cannot be set to LINK in GUI. Set address 0x8D0 in slot 0 to set 1 for trigger modules. This needs to be fixed in Driver by Tim. | '''Note:''' Clock Source cannot be set to LINK in GUI. Set address 0x8D0 in slot 0 to set 1 for trigger modules. This needs to be fixed in Driver by Tim. | ||
Latest revision as of 21:27, September 20, 2023
start DAQ
There are 3 controls associated with Run Control on the main page of either GUI
- Start/Stop - this is self explanatory. Pushing to start begins taking data.
- NoSave/Save - If set to Save, IOC buffers fill with data.
To start the DGS and DFMA systems you usually do the following:
- on the DGS side, find the window where the DGS data is stored. There will be a start.sh script there. Start the receivers up by typing
start.sh <runNo>
where <runNo> is the run number you want. A bunch of windows should show up, one for each receiver/IOC pair
- On the DFMA side, find the where the DFMA data is stored. There will be a start.sh script there. Start the receivers up by typing
start.sh <runNo>
where <runNo> is the run number you want, the same as what you used on the DGS side please. A bunch of windows should show up, one for each receiver/IOC pair now in the FMA setup
- If necessary, synch the clocks of the two systems by clicking on the Impeartive sync button on the DGS side. Please don't click on the Impeartive sync button on the DFMA side or bad things will happen...
- now, as simultaneous as you can, click the start buttons on the DGS and DFMA DAQ windows. You shoImpeartiveuld see the receivers on both systems receiving data.
- may want to check that the data files are growing to be sure data is coming in
- check that the dssd_liveTS show the same numbers on both sides (FMA-GS-Sync) on Darek computer
If not : click on imp sync (GS side) but this should not happen very often
- Check the digitizers counters at the FMA show the same
stop runs in both sides : click on stop in the commander windows and then stop the receivers : type Ctrl C on all the receivers (IOC1, OIC2 ....)
numbers on the left side of the window (AcEvCnt- AcHtCnt-DsCnt) and all dizitizers (right 3 raws of the window)
- Check the MaterDiag counters : should show numbers if the beam is on
- Check Sender Summary (VME SummaryDFMA window), the IOC's should keep updating at DGS
stop DAQ
stop the run on both sides : click on stop in the commander windows
and then stop the receivers : type Ctrl-C on all the receivers (IOC1, OIC2 ....)
Linking two or more systems
write your instructions here
syncing two systems
For DGS
- Goto trig0 window.
- Turn off SYNC U - listed under LRU Control
- Make sure receive/send power is on for L & U
Note: If SYNC U is turned off - an Impeartive sync by DGS will reset the master clock on DFMA.
For DFMA
- Goto Trig0 Window
- Under Misc Control - turn on both undefined buttons
- Pull up diagnotsitc window on trig0.
- Set Clrk Source to LINK
- Set Nim1 and Nim2 out on master trigger module FMA
Note: Clock Source cannot be set to LINK in GUI. Set address 0x8D0 in slot 0 to set 1 for trigger modules. This needs to be fixed in Driver by Tim.