The Pickoff Card: Difference between revisions
No edit summary |
No edit summary |
||
Line 2: | Line 2: | ||
The large 50-conductor grey cables run between the Slope Box on each detector to the Pickoff cards in the VXI crates. | The large 50-conductor grey cables run between the Slope Box on each detector to the Pickoff cards in the VXI crates. | ||
[[File:VXI Pickoff | [[File:VXI Pickoff card top view.jpg]] | ||
The VXI Pickoff Card receives the single-ended signals for the Ge Center, Ge Sides, and BGO segments from the Slope Box. The VXI pickoff converts these to differential signal format before sending them to the digitizer. | The VXI Pickoff Card receives the single-ended signals for the Ge Center, Ge Sides, and BGO segments from the Slope Box. The VXI pickoff converts these to differential signal format before sending them to the digitizer. | ||
There are three gain range options (jumper selectable) for the Ge Center: 2, 5, 10 MeV | |||
[[File:VXI Pickoff cards Bottom.jpg]] | |||
===VXI Pickoff Card Features=== | |||
There are three gain range options (jumper selectable from the bottom of the pickoff card) for the Ge Center: 2, 5, 10 MeV | |||
There is a fixed Preamp Reset Clamp that drives the Ge Center back to baseline after a preamp reset. This sometimes has known oscillation problems (ringing) that can hinder resolution. | There is a fixed Preamp Reset Clamp that drives the Ge Center back to baseline after a preamp reset. This sometimes has known oscillation problems (ringing) that can hinder resolution. | ||
There is an analog output on the VXI board. This LEMO output is shaped similar to a NIM Spectroscopy Amplifier, it is NOT what gets sent to the digitizer. |
Revision as of 18:19, September 20, 2021
The VXI Pickoff Card
The large 50-conductor grey cables run between the Slope Box on each detector to the Pickoff cards in the VXI crates.
The VXI Pickoff Card receives the single-ended signals for the Ge Center, Ge Sides, and BGO segments from the Slope Box. The VXI pickoff converts these to differential signal format before sending them to the digitizer.
VXI Pickoff Card Features
There are three gain range options (jumper selectable from the bottom of the pickoff card) for the Ge Center: 2, 5, 10 MeV
There is a fixed Preamp Reset Clamp that drives the Ge Center back to baseline after a preamp reset. This sometimes has known oscillation problems (ringing) that can hinder resolution.
There is an analog output on the VXI board. This LEMO output is shaped similar to a NIM Spectroscopy Amplifier, it is NOT what gets sent to the digitizer.