1) Verify optic alignment
Perform CCD changeover to Coolsnap
Drive focusing optic into beam, adjust OSAX and OSAY so outgoing focused beam is not clipped and remnant central parallel beam intensity is minimized
Close NES slits to 0.1x0.1, adjust FOMX and FOMY to center optic on beam axis - outgoing wavefront clipped by slits should look symmetric around the edges

NOTES:
i) Central stop may be offset from the center of optic - do not simply put image of central stop in the center of the beam footprint

2) Mount sample
Perform CCD changeover to Coolsnap
Drive focusing optic into beam, adjust OSAX and OSAY so outgoing focused beam is not clipped and remnant central parallel beam intensity is minimized
Close NES slits to 0.1x0.1, adjust FOMX and FOMY to center optic on beam axis - outgoing wavefront clipped by slits should look symmetric around the edges

NOTES:
i) Central stop may be offset from the center of optic - do not simply put image of central stop in the center of the beam footprint