1) Scan mirror piezo:
Set slits to

Slit H Size (mm) V Size (mm)
PBS 0.2 2
NES 2 2
BDA 2 -

scan mirror piezo +/- 1 V. Set mirror piezo to the maximum.
NOTES: i) mirror piezo feedback should be OFF
ii) CCD shutter should be set to "manual open"

2) Scan DCM roll:
Set slits to

Slit H Size (um) V Size (um)
PBS 0.2 2
NES 0.2 2
BDA 0.025 -

and scan roll +/- 0.25 mrad. Set roll to maximum, but correct for back lash.


3) Scan DCM pitch:
Set slits to

Slit H Size (um) V Size (um)
PBS 2 2
NES 2 2
BDA 0.025 -

and scan pitch +/- 0.05 mrad. Set pitch to max.


4) Scan DCM Bragg: