Beamline Alignment Quick Reference
Jump to navigation
Jump to search
Back to X-Ray Microscopy
1) Scan mirror piezo:
Set slits to
Slit | H Size (mm) | V Size (mm) |
PBS | 0.2 | 2 |
NES | 2 | 2 |
BDA | 2 | - |
scan mirror piezo +/- 1 V. Set mirror piezo to the maximum.
NOTES: i) mirror piezo feedback should be OFF
ii) CCD shutter should be set to "manual open"
2) Scan DCM roll:
Set slits to
Slit | H Size (um) | V Size (um) |
PBS | 0.2 | 2 |
NES | 0.2 | 2 |
BDA | 0.025 | - |
and scan roll +/- 0.25 mrad. Set roll to maximum, but correct for back lash.
3) Scan DCM pitch:
Set slits to
Slit | H Size (um) | V Size (um) |
PBS | 2 | 2 |
NES | 2 | 2 |
BDA | 0.025 | - |
and scan pitch +/- 0.05 mrad. Set pitch to max.
4) Scan DCM Bragg: