Wiring of Ion Chambers: Difference between revisions
Jump to navigation
Jump to search
No edit summary |
No edit summary |
||
Line 29: | Line 29: | ||
* (b) From VtF converter to 32-ch Scaler, CH #3, IOC26IDSCLR | * (b) From VtF converter to 32-ch Scaler, CH #3, IOC26IDSCLR | ||
[Maser, Apr/18/09] | |||
[[Category:XMG]][[Category:Beamline]] | |||
[ |
Latest revision as of 00:48, April 19, 2009
Back to X-Ray Microscopy
Type: S-1329A, Oken, Tokyo
Spec: Electrode length: 33 cm
Applied Bias: - 2.5 kV
http://www.oken.co.jp/h/pdl/eng/ics/i1.html
Name: Upstream Ion Chamber, ICus
Purpose: measure incident photon flux downstream of MBS
Wiring:
- 26IDC patch panel (CH-3): output Stanford preamp to patch panel "CH-3", connector "#57 Control-2"
- Controls room patch panel "Control-2", connector "#57 (CH-3)" to VtF converter
- (a) From VtF converter to NIM/TTL converter
- from NIM/TTL converter to patch panel "Controls-4", connector "#801"
- from "Controls-4 #801" to Joerger Scaler Ch #1
- (b) From VtF converter to 32-ch Scaler, CH #2, IOC26IDSCLR
Name: Downstream Ion Chamber, ICds
Purpose: measure incident photon flux downstream of sample and imaging optics, upstream of area detectors
Wiring:
- 26IDC patch panel (CH-3): output Stanford preamp to patch panel "CH-3", connector "#58 Control-2"
- Controls room patch panel "Control-2", connector "#58 (CH-3)" to VtF converter
- (a) From VtF converter to NIM/TTL converter
- from NIM/TTL converter to patch panel "Controls-4", connector "#802"
- from "Controls-4 #801" to Joerger Scaler Ch #2
- (b) From VtF converter to 32-ch Scaler, CH #3, IOC26IDSCLR
[Maser, Apr/18/09]