New Huber slits are not working: Difference between revisions

From CNM Wiki
Jump to navigation Jump to search
(New page: New slits have to be calibrated first. '''Solution:'''<br /> 1. Manually close slits to zero position<br /> 2. In spec, run the command “mot_par(motor,”calibrate”). One can also us...)
 
No edit summary
Line 1: Line 1:
Back to [[X-Ray Microscopy]]
New slits have to be calibrated first.
New slits have to be calibrated first.



Revision as of 18:15, April 14, 2009

Back to X-Ray Microscopy

New slits have to be calibrated first.


Solution:
1. Manually close slits to zero position
2. In spec, run the command “mot_par(motor,”calibrate”). One can also use “help xiahsc” to find useful information in spec
3. Type the unique serial numbers of the slits in the field 'generic parameters' of the spec configuration file