'''4) Scan DCM Bragg:'''
'''4) Scan DCM Bragg:'''
[[Category:XMG]][[Category:Controls]][[Category:Beamline]]

Revision as of 17:29, April 14, 2009

Back to X-Ray Microscopy

1) Scan mirror piezo:
Set slits to

Slit H Size (um) V Size (um)
PBS 2 2
NES 2 2
BDA 0.025 -

and scan mirror piezo +/- 1 V. Set mirror piezo to the maximum.


2) Scan DCM roll:
Set slits to

Slit H Size (um) V Size (um)
PBS 0.2 2
NES 0.2 2
BDA 0.025 -

and scan roll +/- 0.25 mrad. Set roll to maximum, but correct for back lash.


3) Scan DCM pitch:
Set slits to

Slit H Size (um) V Size (um)
PBS 2 2
NES 2 2
BDA 0.025 -

and scan pitch +/- 0.05 mrad. Set pitch to max.


4) Scan DCM Bragg: