Beamline Alignment Quick Reference: Difference between revisions

From CNM Wiki
Jump to navigation Jump to search
(New page: '''1) Scan mirror piezo:'''<br /> Set slits to {| {{table}} style="text-align:center" | align="center" style="background:#f0f0f0;"|'''Slit''' | align="center" style="background:#f0f0f0;"|'...)
 
No edit summary
Line 1: Line 1:
Back to [[X-Ray Microscopy]]
'''1) Scan mirror piezo:'''<br />
'''1) Scan mirror piezo:'''<br />
Set slits to
Set slits to

Revision as of 17:28, April 14, 2009

Back to X-Ray Microscopy

1) Scan mirror piezo:
Set slits to

Slit H Size (um) V Size (um)
PBS 2 2
NES 2 2
BDA 0.025 -

and scan mirror piezo +/- 1 V. Set mirror piezo to the maximum.


2) Scan DCM roll:
Set slits to

Slit H Size (um) V Size (um)
PBS 0.2 2
NES 0.2 2
BDA 0.025 -

and scan roll +/- 0.25 mrad. Set roll to maximum, but correct for back lash.


3) Scan DCM pitch:
Set slits to

Slit H Size (um) V Size (um)
PBS 2 2
NES 2 2
BDA 0.025 -

and scan pitch +/- 0.05 mrad. Set pitch to max.


4) Scan DCM Bragg: