Beamline Alignment Quick Reference: Difference between revisions
Jump to navigation
Jump to search
(New page: '''1) Scan mirror piezo:'''<br /> Set slits to {| {{table}} style="text-align:center" | align="center" style="background:#f0f0f0;"|'''Slit''' | align="center" style="background:#f0f0f0;"|'...) |
No edit summary |
||
Line 1: | Line 1: | ||
Back to [[X-Ray Microscopy]] | |||
'''1) Scan mirror piezo:'''<br /> | '''1) Scan mirror piezo:'''<br /> | ||
Set slits to | Set slits to |
Revision as of 17:28, April 14, 2009
Back to X-Ray Microscopy
1) Scan mirror piezo:
Set slits to
Slit | H Size (um) | V Size (um) |
PBS | 2 | 2 |
NES | 2 | 2 |
BDA | 0.025 | - |
and scan mirror piezo +/- 1 V. Set mirror piezo to the maximum.
2) Scan DCM roll:
Set slits to
Slit | H Size (um) | V Size (um) |
PBS | 0.2 | 2 |
NES | 0.2 | 2 |
BDA | 0.025 | - |
and scan roll +/- 0.25 mrad. Set roll to maximum, but correct for back lash.
3) Scan DCM pitch:
Set slits to
Slit | H Size (um) | V Size (um) |
PBS | 2 | 2 |
NES | 2 | 2 |
BDA | 0.025 | - |
and scan pitch +/- 0.05 mrad. Set pitch to max.
4) Scan DCM Bragg: