Beamline Alignment Quick Reference: Difference between revisions
No edit summary |
No edit summary |
||
Line 1: | Line 1: | ||
Back to [[X-Ray Microscopy]] | Back to [[X-Ray Microscopy]] | ||
'''1) Select x-ray energy:'''<br /> | |||
Calculate scattering angle of Si 111 for desired energy with Bragg law calculator <br /> | |||
Drive DCM Theta to the correct Bragg angle for the desired energy, drive undulators approximately to this energy as well <br /> | |||
NOTES:<br /> | |||
i) If DCM theta looks low (~2-4 deg) you may need to re-set user offset (current calibrated theta offset = dial + 8.646deg) <br /> | |||
ii) Approximate Si 111 Bragg Theta = asin(1.977084/E keV)<br /> | |||
'''1) Scan mirror piezo:'''<br /> | '''1) Scan mirror piezo:'''<br /> | ||
Line 16: | Line 25: | ||
|} | |} | ||
scan mirror piezo +/- 1 V. Set mirror piezo to the maximum.<br /> | scan mirror piezo +/- 1 V. Set mirror piezo to the maximum.<br /> | ||
NOTES: i) mirror piezo feedback should be OFF <br /> | NOTES:<br /> | ||
i) mirror piezo feedback should be OFF <br /> | |||
ii) CCD shutter should be set to "manual open"<br /> | ii) CCD shutter should be set to "manual open"<br /> | ||
Revision as of 18:45, September 7, 2010
Back to X-Ray Microscopy
1) Select x-ray energy:
Calculate scattering angle of Si 111 for desired energy with Bragg law calculator
Drive DCM Theta to the correct Bragg angle for the desired energy, drive undulators approximately to this energy as well
NOTES:
i) If DCM theta looks low (~2-4 deg) you may need to re-set user offset (current calibrated theta offset = dial + 8.646deg)
ii) Approximate Si 111 Bragg Theta = asin(1.977084/E keV)
1) Scan mirror piezo:
Set slits to
Slit | H Size (mm) | V Size (mm) |
PBS | 0.2 | 2 |
NES | 2 | 2 |
BDA | 2 | - |
scan mirror piezo +/- 1 V. Set mirror piezo to the maximum.
NOTES:
i) mirror piezo feedback should be OFF
ii) CCD shutter should be set to "manual open"
2) Scan DCM roll:
Set slits to
Slit | H Size (um) | V Size (um) |
PBS | 0.2 | 2 |
NES | 0.2 | 2 |
BDA | 0.025 | - |
and scan roll +/- 0.25 mrad. Set roll to maximum, but correct for back lash.
3) Scan DCM pitch:
Set slits to
Slit | H Size (um) | V Size (um) |
PBS | 2 | 2 |
NES | 2 | 2 |
BDA | 0.025 | - |
and scan pitch +/- 0.05 mrad. Set pitch to max.
4) Scan DCM Bragg: