Line 8: | Line 8: | ||
2. In spec, run the command “mot_par(motor,”calibrate”). One can also use “help xiahsc” to find useful information in spec<br /> | 2. In spec, run the command “mot_par(motor,”calibrate”). One can also use “help xiahsc” to find useful information in spec<br /> | ||
3. Type the unique serial numbers of the slits in the field 'generic parameters' of the spec configuration file | 3. Type the unique serial numbers of the slits in the field 'generic parameters' of the spec configuration file | ||
[[Category:MLL]] |
Latest revision as of 18:16, April 14, 2009
Back to X-Ray Microscopy
New slits have to be calibrated first.
Solution:
1. Manually close slits to zero position
2. In spec, run the command “mot_par(motor,”calibrate”). One can also use “help xiahsc” to find useful information in spec
3. Type the unique serial numbers of the slits in the field 'generic parameters' of the spec configuration file