Beamline Alignment Quick Reference: Difference between revisions
Jump to navigation
Jump to search
No edit summary |
No edit summary |
||
Line 5: | Line 5: | ||
{| {{table}} style="text-align:center" | {| {{table}} style="text-align:center" | ||
| align="center" style="background:#f0f0f0;"|'''Slit''' | | align="center" style="background:#f0f0f0;"|'''Slit''' | ||
| align="center" style="background:#f0f0f0;"|'''H Size ( | | align="center" style="background:#f0f0f0;"|'''H Size (mm)''' | ||
| align="center" style="background:#f0f0f0;"|'''V Size ( | | align="center" style="background:#f0f0f0;"|'''V Size (mm)''' | ||
|- | |- | ||
|- | |- | ||
| PBS || | | PBS || 0.2 || 2 | ||
|- | |- | ||
| NES || 2 || 2 | | NES || 2 || 2 | ||
|- | |- | ||
| BDA || | | BDA || 2 ||| - | ||
|} | |} | ||
scan mirror piezo +/- 1 V. Set mirror piezo to the maximum.<br /> | |||
NOTES: i) mirror piezo feedback should be OFF <br /> | |||
ii) CCD shutter should be set to "manual open"<br /> | |||
'''2) Scan DCM roll:'''<br /> | '''2) Scan DCM roll:'''<br /> |
Revision as of 16:19, September 7, 2010
Back to X-Ray Microscopy
1) Scan mirror piezo:
Set slits to
Slit | H Size (mm) | V Size (mm) |
PBS | 0.2 | 2 |
NES | 2 | 2 |
BDA | 2 | - |
scan mirror piezo +/- 1 V. Set mirror piezo to the maximum.
NOTES: i) mirror piezo feedback should be OFF
ii) CCD shutter should be set to "manual open"
2) Scan DCM roll:
Set slits to
Slit | H Size (um) | V Size (um) |
PBS | 0.2 | 2 |
NES | 0.2 | 2 |
BDA | 0.025 | - |
and scan roll +/- 0.25 mrad. Set roll to maximum, but correct for back lash.
3) Scan DCM pitch:
Set slits to
Slit | H Size (um) | V Size (um) |
PBS | 2 | 2 |
NES | 2 | 2 |
BDA | 0.025 | - |
and scan pitch +/- 0.05 mrad. Set pitch to max.
4) Scan DCM Bragg: