Beamline Alignment Quick Reference: Difference between revisions

From CNM Wiki
Jump to navigation Jump to search
No edit summary
No edit summary
Line 35: Line 35:
{| {{table}} style="text-align:center"
{| {{table}} style="text-align:center"
| align="center" style="background:#f0f0f0;"|'''Slit'''
| align="center" style="background:#f0f0f0;"|'''Slit'''
| align="center" style="background:#f0f0f0;"|'''H Size (um)'''
| align="center" style="background:#f0f0f0;"|'''H Size (mm)'''
| align="center" style="background:#f0f0f0;"|'''V Size (um)'''
| align="center" style="background:#f0f0f0;"|'''V Size (mm)'''
|-
|-
|-
|-
Line 48: Line 48:
NOTES:<br />
NOTES:<br />
i) Correct for backlash by driving to first scan point, then to maximum (e.g. after scan drive relative -0.15, then drive abs to max)<br />
i) Correct for backlash by driving to first scan point, then to maximum (e.g. after scan drive relative -0.15, then drive abs to max)<br />
ii) Settling time of at least 1 second (this axis is UHV picomotor driven in PID)
ii) Settling time of at least 1 second is necessary for this scan (this axis is UHV picomotor driven in PID)




'''3) Scan DCM pitch:'''<br />
'''4) Scan DCM pitch:'''<br />
Set slits to  
Set slits to  
{| {{table}} style="text-align:center"
{| {{table}} style="text-align:center"
| align="center" style="background:#f0f0f0;"|'''Slit'''
| align="center" style="background:#f0f0f0;"|'''Slit'''
| align="center" style="background:#f0f0f0;"|'''H Size (um)'''
| align="center" style="background:#f0f0f0;"|'''H Size (mm)'''
| align="center" style="background:#f0f0f0;"|'''V Size (um)'''
| align="center" style="background:#f0f0f0;"|'''V Size (mm)'''
|-
|-
|-
|-
Line 63: Line 63:
|    NES ||    2    ||  2           
|    NES ||    2    ||  2           
|-  
|-  
|    BDA || 0.025  |||  -             
|    BDA ||   2    |||  -             
|}
|}
and scan pitch +/- 0.05 mrad. Set pitch to max.
and scan pitch +/- 0.05 mrad. Set pitch to max.

Revision as of 18:56, September 7, 2010

Back to X-Ray Microscopy

1) Select x-ray energy:
Calculate scattering angle of Si 111 for desired energy with Bragg law calculator
Drive DCM Theta to the correct Bragg angle for the desired energy, drive undulators approximately to this energy as well

NOTES:
i) If DCM theta looks low for the current energy (~2-4 deg) you may need to re-set user offset (current calibrated theta offset = dial + 8.646deg)
ii) Approximate Si 111 Bragg Theta = asin(1.977084/E keV)
iii) Approximate undulator offsets are ~ desired energy +0.1 keV


2) Scan mirror piezo:
Set slits to

Slit H Size (mm) V Size (mm)
PBS 0.2 2
NES 2 2
BDA 2 -

scan mirror piezo +/- 1 V. Set mirror piezo to the maximum.
NOTES:
i) mirror piezo feedback should be OFF
ii) CCD shutter should be set to "manual open"

3) Scan DCM roll:
Set slits to

Slit H Size (mm) V Size (mm)
PBS 0.2 2
NES 0.2 2
BDA 2 -

scan roll +/- 0.15 mrad. Set roll to maximum
NOTES:
i) Correct for backlash by driving to first scan point, then to maximum (e.g. after scan drive relative -0.15, then drive abs to max)
ii) Settling time of at least 1 second is necessary for this scan (this axis is UHV picomotor driven in PID)


4) Scan DCM pitch:
Set slits to

Slit H Size (mm) V Size (mm)
PBS 2 2
NES 2 2
BDA 2 -

and scan pitch +/- 0.05 mrad. Set pitch to max.


4) Scan DCM Bragg: